The special features:
- Completely stainless steel piping
- Process gas recirculation
- IP 65 stainless steel housing
- High-precision infrared measurement technology for CO, CO2, CH4
- Thermal conductivity technology for H2
- Paramagnetic O2 measurement
- Gas sampling in the pressure range from -100 mbar up to +200 mbar
- Integrated gas cooler with maintenance-free condensate drainage
- Measurement of up to 6 gas sampling points
Areas of application
- Biomass and waste gasification
- Coal gasification, coal liquefaction, synthetic fuel
- Pyrolysis applications and steam reforming
- Syngas analysis in laboratories and research
- IP65 - Stainless steel housing (AISI316Ti), 600 x 700 x 210 mm (W x H x D)
- All piping and fittings inside the analyzer made of stainless steel
- Including paramagnetic O2 measuring cell 0 - 25.00%
- Including NDIR for CO - CO2 - CH4 analysis (with selective CH4 measurement), 0...100% CO / 0...100% CO2 / 0...100% CH4
- Including H2 - thermal conductivity detector - measurement 0...100%
- Electric sample gas cooler (Peltier type) and automatic condensate drain via controlled solenoid valve
- Regulated sample gas pump with sample gas flow monitoring
- Internal sample gas flow sensor
- Solenoid valve for automatic zero adjustment (with clean ambient air or nitrogen) and for manual calibration
- 1x calibration gas inlet
- 1x stainless steel sample gas inlet from -150 mbar to +200 mbar - 1/8" internal thread - incl. flame arrester
- 1x outlet for process gas recirculation made of stainless steel - ¼" internal thread - incl. flame arrestor
- 1x zero gas outlet
- Interruption of the sample gas supply in the event of a system alarm by a "fail-safe" solenoid valve (normally closed)
- Inlet for connecting nitrogen to purge the syngas-carrying parts to prevent an explosive mixture inside
- External box for power supply, for automatic power interruption in the event of danger
- Gas detector (%LEL CH4) for continuous monitoring of the air in the analyzer housing
- Operation via TFT color display with membrane keypad and RS485 interface (Modbus RTU)
- Intuitive menus
- 110...230 Vac / 47 - 63 Hz / 200 W power consumption
CO | 0 ... 100% (NDIR) |
CO2 | 0 ... 100% (NDIR) |
CH4 | 0 ... 100% (NDIR) |
O2 | 0 ... 25% (EC) |
O2 | 0 ... 25% (paramagnetic) |
H2S | 0 ... 2.000/5.000 ppm (EC) |
H2 | 0 ... 1,00%/100% TCD |
Weight | 45 kg |
Power supply | Universal 90-240 Vac / 47-63 Hz / 200W |
Dimensions | 700 x 600 x 210 mm (H x W x D) for wall or rack mounting |
Gas conditioning | Electric gas cooler (Peltier) with constant dew point +5° C |
Interfaces | 4 x analog output 4-20 mA, floating, max. load 500R 4 x analog input 4-20 mA, passive inputs RS485 digital interface (Modbus RTU) |
Protection | Sample gas shut-down solenoid valve Power supply cut-off in case of system alarm |